participation in
11th International Intercomparison of
Environmental Dosimeters 1996/97
- Environmental Measurements Laboratory (EML) of the US
Department of Energy (DOE), New York/ USA
in collaboration with
- Brookhaven National Laboratory (BNL), Upton, NY/ USA
- National Institute of Standards (NIST), Gaithersburg, MD,/
USA
- Idaho State University, Pocatello, ID/ USA
|
The organizers were positively surprised of the performance
with latest progress in electronic environmental dosimetry. Mrs.
Gladys Klemic *) agreed to be
at disposal for any further questions. |
|
GammaTRACER is in concordance with in the target values´
error at all tests. |
|
The deviation of the GammaTRACER relative to the Reuter
Stokes reference instrumentation is less than both instruments´
error bars. |
|
|
*) Address: U.S. Department of
Energy, Environmental Measurement Lab., Fax: ++1 201 620 3600,
e-mail:
klemic@eml.doe.gov |
Results and general information
Time of execution:
May - September 1996
Participants:
121 participants from 31 countries
Dosimeter types:
|
152
|
sets
|
Thermoluminescent dosimeters (TLDs)
|
|
3
|
"
|
Film dosimeters
|
|
5
|
"
|
Electrets
|
|
2
|
"
|
Phosphate glass
|
|
1
|
"
|
Geiger-Müller device
|
Experiments and references:
|
Labelling
|
Exposure
|
Reference units
|
|
FIELD
|
outdoors for ca. 90 days at BNL
|
Pressurized ionization chamber
(PIC)
|
|
LAB Am
|
to 241Am source in laboratory for a few hours
|
NIST ionization chamber with pancake probe
|
|
LAB Cs
|
to 137 Cs source in laboratory
|
Shonka-type ionization chamber
(NIST-calibrated)
|
|
CONTROL
|
stored in a steel vault over the whole time for recording
the dose for compensation.
|
Pressurized ionization chamber
(PIC)
|
|