GammaTRACER
Basic: Technical Data
Detector |
Measuring properties |
Data storage |
Data transmission |
Power supply | Mechanical
specification | Options
|
Type:
|
2 x Geiger-Müller-tube
|
|
Position:
|
Centre of casing, paralled to longitudinal axis
|
|
Sensitivity:
|
2 x 150 impulses in 10 minutes at 100 nSv/h
|
|
Measuring range:
|
10 nSv/h - 10 mSv/h
|
|
1-sigma-error:
|
+/- 3 % (at 100 nSv/h with 60-min measuring cycle)
|
|
Calibration error:
|
+/- 6 % calibrated with 137 Cs
|
|
Calibration radiation:
|
662 keV (Cs-137)
|
|
Energy dependence:
|
+/- 40% (over range 45 to 3000keV) |
|
Quantities measured:
|
Air kerma (nGy/h), exposure rate (mR/h), photon dose
equivalent rate (nSv/h)
|
|
Measuring cycle:
|
1, 2, 5, 10, 15, 30, 60, 120 min (user selectable)
|
|
Data capacity:
|
12.800 measurement points (4 - 1.060 days)
|
|
Maxim.archive time:
|
8.8 days (1 minute cycle) to 2 years 11 months (2 hour
cycle)
|
|
Infrared:
|
up to 10 m
|
| RS 232 interface |
up to 10 m (1.000 V galvanically isolated) |
|
RS-485 interface
|
up to 500 m, more with repeater (1.000 V galvanically isolated) |
| Radio module SkyLINK |
up to 100 km (60 miles) @ 10 mW |
| Radio module ShortLINK |
Up to 5 km (3 miles) @ 10 mW |
| GSM / ISDN / modem (via DataGATE) |
worldwide |
| ASC II interface |
Simple ASC II interface for system integration |
|
Operating time at a measurement cycle of 60 min
(at 200 nSv/h) |
typically 5 years, guaranteed 3 years,
typically 10 years with battery extension pack |
|
Ext. power supply |
12 V / 500 mW (only with RS485 option) |
|
Measures ([emptyset]/L):
|
60 mm diameter x 690 mm length
|
|
Casing material:
|
anodised aluminum, wall thickness 1 mm
|
|
Weight:
|
950 g
|
|
Protection mode:
|
hermetically sealed
(Cable versions: IP67; shorttime plunging, max. 10 sec. up to 1 bar pressure)
|
|
Temperature range:
|
-20 ... +50 °C (-4 ... +140 °F)
|
|
Measuring range:
|
10 nSv/h - 10 Sv/h
|
|
Temperature range:
|
-40 ... +60°C (-104°F ... +158°F)
|
|